Section 246-240-617. Records of periodic spot-checks for teletherapy units.  


Latest version.
  • (1) A licensee shall retain a record of each periodic spot-check for teletherapy units required by WAC 246-240-378 for three years.
    (2) The record must include:
    (a) The date of the spot-check;
    (b) The manufacturer's name, model number, and serial number of the teletherapy unit, source and instrument used to measure the output of the teletherapy unit;
    (c) An assessment of timer linearity and constancy;
    (d) The calculated on-off error;
    (e) A determination of the coincidence of the radiation field and the field indicated by the light beam localizing device;
    (f) The determined accuracy of each distance measuring and localization device;
    (g) The difference between the anticipated output and the measured output;
    (h) Notations indicating the operability of each entrance door electrical interlock, each electrical or mechanical stop, each source exposure indicator light, and the viewing and intercom system and doors; and
    (i) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.
    (3) A licensee shall retain a copy of the procedures required by WAC 246-240-378(2) until the licensee no longer possesses the teletherapy unit.
    [Statutory Authority: RCW 70.98.050. WSR 06-05-019, § 246-240-617, filed 2/6/06, effective 3/9/06.]
RCW 70.98.050. WSR 06-05-019, § 246-240-617, filed 2/6/06, effective 3/9/06.

Rules

246-240-378,246-240-378,