Washington Administrative Code (Last Updated: November 23, 2016) |
Title 246. Health, Department of |
Chapter 246-240. Radiation protection—Medical use of radioactive material. |
Section 246-240-614. Records of teletherapy, remote afterloader, and gamma stereotactic radiosurgery full calibrations.
Latest version.
- (1) A licensee shall maintain a record of the teletherapy unit, remote afterloader unit, and gamma stereotactic radiosurgery unit full calibrations required by WAC 246-240-369, 246-240-372, and 246-240-375 for three years.(2) The record must include:(a) The date of the calibration;(b) The manufacturer's name, model number, and serial number of the teletherapy, remote afterloader, and gamma stereotactic radiosurgery unit(s), the source(s), and the instruments used to calibrate the unit(s);(c) The results and an assessment of the full calibrations;(d) The results of the autoradiograph required for low dose-rate remote afterloader units; and(e) The signature of the authorized medical physicist who performed the full calibration.[Statutory Authority: RCW 70.98.050. WSR 06-05-019, § 246-240-614, filed 2/6/06, effective 3/9/06.]
RCW 70.98.050. WSR 06-05-019, § 246-240-614, filed 2/6/06, effective 3/9/06.
Rules
246-240-369,246-240-372,246-240-375,